Título: Precisión de la conductometría con el localizador apical electrónico Propex pixi en dientes premolares in vitro, 2020
Asesor(es): Sologuren Anchante, Rafael Fernando
Campo OCDE: http://purl.org/pe-repo/ocde/ford#3.02.14
Fecha de publicación: 2021
Institución: Universidad Nacional de la Amazonía Peruana
Resumen: El propósito del presente estudio fue establecer la precisión de la conductometria con el localizador apical electrónico Propex Pixi en dientes premolares in vitro, 2020. El tipo de investigación fue cuantitativa, el diseño cuasi-experimental, descriptivo comparativo, transversal. La muestra estuvo conformada por 20 dientes premolares. Los resultados más destacados fueron: Hubo 02 coincidencias en las medidas de conductimetrias; las diferencias fluctuaron entre 0,5mm y 1,5mm. En 17 piezas dentales (85%) la diferencia es menor a 0. El 10 % de las medidas coincide la conductometria con Propex Pixi y Anatómica.Se encontraron diferencias significativas entre las conductometrias con Propex Pixi y anatómica (p=0,000), por lo que concluimos que la conductometria con el localizador apical electrónico no puede realizarse como una única opción, sino que tiene mucha importancia y siempre debe ser tomado en cuenta también, la conductometria convencional. La longitud más frecuente fue de 21 mm con el 20 %, seguido de 20 mm y 22 mm con el 15 % cada medida. El 20 % presentó una conductometria de 20 mm, el 15 % presento 18 mm en la conductometria de premolares con el Propex Pixi

The purpose of the present study was to establish the precision of conductometry with the Propex Pixi electronic apical locator in premolar teeth in vitro, 2020. The type of research was quantitative, the design was quasi-experimental, descriptive, comparative, cross-sectional. The sample consisted of 20 premolar teeth. The most outstanding results were: There were 02 coincidences in the conductimetry measurements; the differences ranged between 0.5mm and 1.5mm. In 17 teeth (85%) the difference is less than 0. 10% of the measurements coincide with the conductometry with Propex Pixi and Anatomical. Significant differences were found between the conductometries with Propex Pixi and anatomical (p = 0.000), therefore We conclude that conductometry with the electronic apical locator cannot be performed as a single option, but is very important and should always be taken into account as well, conventional conductometry. The most frequent length was 21 mm with 20%, followed by 20 mm and 22 mm with 15% each measurement. 20% presented a conductometry of 20 mm, 15% presented 18 mm in the conductometry of premolars with the Propex Pixi.
Disciplina académico-profesional: Odontología
Institución que otorga el grado o título: Universidad Nacional de la Amazonía Peruana. Facultad de Odontología
Grado o título: Cirujano Dentista
Jurado: Chávez Paredes, Alejandro; López Alama, Luis Enrique; Carranza Del Águila, Raúl
Fecha de registro: 30-abr-2021



Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons