Title: Suq’uy: desarrollo y validación de un sistema de monitoreo de succión para recién nacidos
Authors(s): Vela Antón, Paulo Camilo Alberto
Advisor(s): Cárcamo Cavagnaro, César Paul Eugenio; Bellomo Montalvo, Sicilia María Emilia
Keywords: Recién Nacido Pretérmino; Succión; Pediatra; Lactancia; Exactitud; Precisión; Presión Intraoral; Monitoreo; Visualización
OCDE field: http://purl.org/pe-repo/ocde/ford#1.02.00; http://purl.org/pe-repo/ocde/ford#3.02.03
Issue Date: 2021
Institution: Universidad Peruana Cayetano Heredia
Abstract: Objetivo: Desarrollar y validar un sistema de monitoreo de parámetros de succión nutritiva no invasivo en recién nacidos del Hospital Nacional Cayetano Heredia. Material y métodos: Estudio descriptivo con desarrollo tecnológico donde se diseñó, manufacturó y validó un sistema de monitoreo de succión para neonatos. Posteriormente, se realizaron tomas de medidas de presión de succión intraoral en 30 recién nacidos a término. A partir de las señales generadas se describieron las características de los patrones encontrados y se definió los parámetros más eficientes para la evaluación cuantitativa de la succión. Resultados: Los principales parámetros clínicos a considerar fueron edad gestacional, peso al nacer y somnolencia. Los parámetros cuantitativos fueron máxima presión de succión, succiones por minuto, ráfagas de succión, entre otros. El sistema desarrollado muestra un error cuadrático medio de 0.05 para medidas de presión intraoral. Se encontraron tres clasificaciones de patrones de succión y cuatro tipos de pulsaciones de succión, así como los rangos de los parámetros cuantitativos de succión en neonatos saludables. Conclusiones: Se han descrito los rangos de los parámetros y los tipos de patrones cuantitativos de la succión nutricia en recién nacidos a término mediante un dispositivo desarrollado y validado. Estos valores pueden ser un estándar que facilitaría la evaluación en recién nacidos pretérmino con perturbaciones en la succión.
Objective: To develop and validate a non-invasive nutritional suction parameter monitoring system in newborns from the Cayetano Heredia Hospital. Material and methods: Descriptive study with technological development where a sucking monitoring system for neonates was designed, manufactured and validated. Subsequently, intraoral suction pressure measurements were taken in 30 full-term newborns. From the generated signals, the characteristics of the patterns found are described and the most efficient parameters for the quantitative evaluation of suction were defined. Results: The main clinical parameters to consider were gestational age, birth weight and sleepiness. The quantitative parameters were maximum suction pressure, suctions per minute, suction bursts, among others. The developed system shows a mean square error of 0.05 for intraoral pressure measurements. Three classifications of suction patterns and four types of suction waves were found, as well as the ranges of the quantitative suction parameters in healthy neonates. Conclusions: The ranges of the parameters and the types of patterns of nutritional sucking in term newborns have been described by means of a developed and validated device. These values can be a standard that would facilitate the assessment of feeding difficulties in preterm newborns with sucking perturbation.
Objective: To develop and validate a non-invasive nutritional suction parameter monitoring system in newborns from the Cayetano Heredia Hospital. Material and methods: Descriptive study with technological development where a sucking monitoring system for neonates was designed, manufactured and validated. Subsequently, intraoral suction pressure measurements were taken in 30 full-term newborns. From the generated signals, the characteristics of the patterns found are described and the most efficient parameters for the quantitative evaluation of suction were defined. Results: The main clinical parameters to consider were gestational age, birth weight and sleepiness. The quantitative parameters were maximum suction pressure, suctions per minute, suction bursts, among others. The developed system shows a mean square error of 0.05 for intraoral pressure measurements. Three classifications of suction patterns and four types of suction waves were found, as well as the ranges of the quantitative suction parameters in healthy neonates. Conclusions: The ranges of the parameters and the types of patterns of nutritional sucking in term newborns have been described by means of a developed and validated device. These values can be a standard that would facilitate the assessment of feeding difficulties in preterm newborns with sucking perturbation.
Link to repository: https://hdl.handle.net/20.500.12866/9697
Discipline: Informática Biomédica en Salud Global con mención en Informática en Salud
Grade or title grantor: Universidad Peruana Cayetano Heredia. Escuela de Posgrado Víctor Alzamora Castro
Grade or title: Maestro en Informática Biomédica en Salud Global con mención en Informática en Salud
Juror: Zegarra Dueñas, Jaime Alfredo Martín; Tenorio Trigoso, Alonso; Altobelli Meier, Laura Catherine